EU alert · Safety Gate Serious risk Protective equipment
Particle filter mask “KN95 Protective Mask 4 Layer Protection KN95” — risk: health risk / other
Published in the Safety Gate weekly report: · Ref. A12/00384/21
Summary
The particle/filter retention of the material is insufficient (measured values: as low as 20 %). Consequently, an excessive amount of particles or microorganisms might pass through the mask, increasing the risk of infection if not combined with additional protective measures. The product does not comply with the PPE Regulation and with the relevant European standard EN 149.
A Safety Gate alert means that national authorities found a serious or relevant risk and took measures. The same product may also be sold under other names or online.
Product
Particle filter mask
- Name / model
- KN95 Protective Mask 4 Layer Protection KN95
- Category
- Protective equipment
- Packed in cardboard box with 20 masks
What is the risk?
- The particle/filter retention of the material is insufficient (measured values: as low as 20 %). Consequently, an excessive amount of particles or microorganisms might pass through the mask, increasing the risk of infection if not combined with additional protective measures. The product does not comply with the PPE Regulation and with the relevant European standard EN 149.
Measures taken
- Type of economic operator to whom the measure(s) were ordered: OtherCategory of measure(s): Recall of the product from end users
- Date of entry into force: 10/02/2021
- Type of economic operator taking notified measure(s): ImporterCategory of measure(s): Recall of the product from end users
- Date of entry into force: Unknown
Photos
Source: Safety Gate — European Commission
Data and images: Safety Gate, © European Union, licensed under CC BY 4.0. EU Safety Radar is an independent information service; the text above is an editorial presentation of the official data — the entry in the source system is authoritative.