Product Recall: Beha-Amprobe MT204-S with Test Leads (2403-0027)
Published by Office for Product Safety and Standards (OPSS):
Reason for the recall
- The product has been assessed as presenting a high risk of electric shock as the inclusion of unshrouded test leads resulted in the machinery tester not being compliant with EN 61577-14 and not compliant with the essential health and safety requirements.
Reasonsee the original notice
Brandnot stated
Batch number(s)not stated
Barcodenot stated
Product
Beha-Amprobe MT204-S with Test Leads
- Name / model
- MT204-S
- Category
- Electrical appliances equipment
- Country of origin
- Slovenia
- Type of recall
- Product recall
- Summary
- Product: Beha-Amprobe MT204-S with Test Leads
- Hazard: The product has been assessed as presenting a high risk of electric shock as the inclusion of unshrouded test leads resulted in the machinery tester not being compliant with EN 61577-14 and not compliant with the essential health and safety requirements.
- Corrective action: Recall of the product from end users.
- Product information
- Type
- Electrical appliances and equipment - Machinery tester with test leads
- Model
- MT204-S
- Affected serial number range
- 4106XXXX – 6211XXXX
- Country of Origin
- Slovenia
- Product Description
- The product is a machinery tester supplied with a set of unshrouded test leads that does not require two-hand operation. The leads are used to interconnect the Machinery Tester to an electrical machine for (high voltage) measurement purposes.
- Product Report and Image
- Link to product image and PDF
- Hazard
- EN 61557-14, requires two-hand operation for test equipment when used for high voltage testing as an additional level of protection against electric shock during use. As of 1 July 2018, the shrouded test leads set initially supplied for use with the Machinery Tester, which required two-hand use, was inadvertently replaced with an alternative unshrouded test leads set that did not require two-hand operation. The inclusion of unshrouded test leads resulted in the machinery tester not being compliant with EN 61577-14 and not compliant with the essential health and safety requirements.
- Corrective action
- The unshrouded product has been recalled and the machinery tester product is now shipped with shrouded test leads.
- Fluke is in the process of informing the distributors of the issue and requesting them (a) to stop selling the Machinery Testers in their stock and (b) to return this stock to Fluke.
- Fluke is also in the process of informing the end-customers of the potential safety risk with instructions to (a) stop using the affected Machinery Testers with the original non-two hands test leads for high voltage testing and (b) register for a free set of shrouded two-hands use test leads for high voltage testing.
- For further information, please access the Beha-Amprobe recall site here: https://www.beha-amprobe.com/en/safety-recallinformation-regarding-amprobe-mt204s-machinerytester
- Additional information
- PSD number: 2403-0027
- Notified by: Health and Safety Executive
What to do
- Recall of the product from end users.
- Product information
- Type
- Electrical appliances and equipment - Machinery tester with test leads
Source: Product Safety Alerts, Reports and Recalls (OPSS) — Office for Product Safety and Standards (OPSS)
Source: Office for Product Safety and Standards, GOV.UK. Contains public sector information licensed under the Open Government Licence v3.0. EU Safety Radar is an independent information service; the text above presents the official data in the original language — the notice published by the authority is authoritative.